Kistler-Morse KM SVS2000 User Manual

Page 27

Advertising
background image

23

www.kistlermorse.com

97-1129-01 Rev. D

groSS WeIghT dISplay

Bar graph dISplay

note: Immediately after Quick Config, the SvS2000 uses the default value for the Bar graph Span to determine
the value for the bar graph display. enter the correct Bar graph Span in the Display menu.

you can choose to view a bar graph of the gross weight. To toggle between numerical and bar graph gross weight:

1. use an Arrow key to scroll to the gross weight display. The display looks like this:

GRS 12345600 lbs

2. Press the enter key. The display looks like this:

75%||||||||| |

Press the Enter Key again to return to the numerical gross weight display.

zero groSS WeIghT dISplay

Zeroing the gross weight quickly adjusts an empty vessel’s calibration by setting the gross weight to zero. Follow
this procedure:

1. use an Arrow key to scroll to the gross weight display. The display looks like this:

GRS 12345600 lbs

2. Press the Zero key. The display shows:

ZerO grS DISPLy?

3. Press the Enter Key to zero the gross weight. Press any other key to abort this function.

4. If you set up a user Access Code the display shows:

USER CODE?

Enter the User Access Code. Press the Enter Key.

5. The display briefly shows:

GROSS ZEROED

notes:

1.

The SvS2000 verifies the gross weight is within user-set tolerance limits, to prevent accidental zeroing

of a vessel that is not empty. If the gross weight is not within tolerance, the display shows Outside Zero Tol and
does not zero the gross weight. See Zero Tolrance in Cal Menu, to set the tolerance limit.

2.

Disp Val Correct can also zero the vessel. However, Disp Val Correct has no tolerance limits and allows

the current gross weight to be set to a non-zero value. See refining the Calibration with the Display value
Correction in Cal Menu.