Preact [31 through 38]
IOut 4/0 mA [42] and 20 mA [43]
DSP Step [73]
rate Threshld [81]
Zero Trak Win [83]
Drift Limit [84]

Example 2. Change Disp Units from lbs to tns and Disp Format from xxxxxx00 to xxxxx.x

record current parameter values:
Disp Format = xxxxxx00
Scale Factor Weight = 100,000 lbs
Relay1 Value = 90,000 lbs
(dummy zeroes underlined)
Change Disp Units and Disp Format.
As a result of the changes, the parameter values change to:
Scale Factor Weight = 100.0 tns
Relay1 Value = 90.0 tns
re-enter parameter values:
Scale Factor Weight =
100,000 lbs ÷ 2000 lbs/ton = 50.0 tns
Relay1 Value =
90,000 lbs ÷ 2000 lbs/ton = 45.0 tns

Example 3. Change Disp Units from lbs to % and Disp Format from xxxxxx00 to xxxxxx

When changing Disp Units to %, you must calculate the current parameter values in terms of the full scale

(100%) load:

new value (in %) = Old value ÷ Full Scale Load where:
Full scale load = vessel working capacity (from Application Data Form)
Current parameter values:
Disp Format = xxxxxx00
Scale Factor Weight = 100,000 lbs
Relay1 Value = 90,000 lbs (dummy zeroes underlined) Change Disp Units and Disp Format. As a result of the

changes, the parameter values change to:

Scale Factor Weight = 1000 %
Relay1 Value = 900 %
The vessel working capacity is 150,000 lbs.
re-enter parameter values:
Scale Factor Weight =
100,000 lbs ÷ 150,000 lbs = 75%
Relay1 Value =
90,000 lbs ÷ 150,000 lbs = 60%