51 SerIal addreS (addreSS)

External equipment may have many signal processors daisy chain wired to it. The serial address of the signal
processor identifies the signal processor and the associated vessel to the external equipment. each device
on the same serial connection must have a different address. The address of the SVS2000’s serial port is set
up in the Serial Addres menu. The address is expressed in two forms— decimal (ranging from 0 to 255) and
hexadecimal (ranging from 0 to FF). The default setting is 255 in decimal form (FF in hexadecimal form). Follow
this procedure to input the address:

1. If the SVS2000 is in Run Mode, press the ‘5’ Key, ‘1’ Key and Enter Key to access the menu tree. The

display shows:

51 SerIAL ADDreS

2. Press the enter key to access this menu. The display looks like this:

>255 DEC FF HEX

The display shows two number fields. The first number is the decimal address. The second number is
the hexadecimal equivalent. Only the decimal form of the address can be altered by the user; the
hexadecimal form changes automatically to correspond to the decimal form.

3. Use the numeric or Arrow Keys to input the decimal address. Press the Enter Key to save the value in

memory. The display acknowledges the entry and returns to:

51 SerIAL ADDreS

4. Press an Arrow Key to scroll to another menu or press the Gross or net Key to return to Run Mode.

52 SerIal Baud

The baud rate (bits per second) is set up in the Serial Baud menu. The baud rate determines the speed with
which the SvS2000 communicates with the PLC, mvS, etc. There are five baud rates available:

1200, 2400, 4800, 9600 (default), 19200

Follow this procedure to select the baud rate:

1. If the SVS2000 is in Run Mode, press the ‘5’ Key, ‘2’ Key, and Enter Key to access the menu tree. The

display shows:

52 SerIAL BAuD

2. Press the enter key to access this menu. The display looks like this:

BAuD rATe:*9600

(Asterisk indicates the current selection.)

If the displayed menu does not have the desired baud rate, press the Up Arrow Key to display the next
page of the menu. Continue to press an Arrow key until you see the desired baud rate (there are five
pages to the Serial Baud menu).

3. When the display shows the desired baud rate, press the Enter Key to save the selection in memory. The

display acknowledges the selection and returns to:

52 SerIAL BAuD

4. Press an Arrow Key to scroll to another menu or press the Gross or net Key to return to Run Mode.