Notice – VACUUBRAND PC 3001 VARIOpro emission condenser Peltronic User Manual

Page 100

Advertising
background image

page 100 of 120

Tools required (metric):
- Diaphragm key width 46 mm

- Open end wrench width 14 / 17 mm

- 4 / 5 mm wide Allen key

- 2.5 mm wide slotted screwdriver

- Phillips screwdriver size 2

- Torx driver T10

- Flat pliers

Service kit MD 1C (diaphragms, valves, diaphragm key) ..............................696828

Ensure that maintenance is done only by suitably

trained and supervised technicians.

The valves and diaphragms are wear parts. If the rated

ultimate vacuum is no longer achieved or in case of in-

creased noise level, the pump interior, the diaphragms

and the valves must be cleaned and the diaphragms and

valves must be checked for cracks or other damage.

All bearings are encapsulated and are filled with long-life

lubricant. Under normal operating conditions, the drive

system is maintenance free.

In demanding circumstances, it may be efficient to check

and clean the pump heads on a regular basis. In normal

use, the lifetime of the diaphragms and valves is typically

15,000 operating hours.

- Prevent internal condensation, transfer of liquids or

dust. The diaphragms and valves will be damaged if

liquid is pumped in significant amount.

- Carry out maintenance frequently if the pump is ex-

posed to corrosive media or in case of deposits.

- Regular maintenance will improve the lifetime of the

pump and also protect both users and the environment.

NOTICE

Advertising
This manual is related to the following products: