Notice – VACUUBRAND PC 3001 VARIOpro emission condenser Peltronic User Manual

Page 16

Advertising
background image

page 16 of 120

• Ensure that no foreign objects can be drawn into the

pump.

• Check the power source and the pump’s rating plate

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

• Ensure that the

coolant outlet pipe is always free and

that it cannot get blocked. If installing an optional cool-

ant valve, it must always be in the supply line of the

exhaust waste vapor condenser.

Make sure ventilation is adequate to maintain recommend-

ed operating temperature. Keep a minimum distance of

2 in (5 cm) between the cooling fan and surrounding items

(e.g., housing, walls, etc.), or else install an external auto-

matic ventilation system. Clean ventilation slots if neces-

sary to avoid a reduction of ventilation.

Do not place the pumping unit on soft surfaces (e.g., rub-

ber foam) during operation. This may reduce or block the

electronics’ air supply. Do not cover the pumping unit.

Use only hoses at the inlet and outlet of the pump with

an inner diameter at least as large as the diameter of the

pump’s tubing (to avoid overpressure at the outlet, and

reduction of pumping speed at the inlet).

Allow the equipment to equilibrate to ambient temperature

if you bring it from cold environment into a room prior to

operation. Notice if there is water condensation on cold

surfaces.

Secure coolant hoses at the hose nozzles (e.g., with hose

clamp) to prevent their accidental slipping.

Comply with all

applicable and relevant safety require-

ments (regulations and guidelines). Implement the re-

quired actions and adopt suitable safety measures.

NOTICE

Advertising
This manual is related to the following products: