Notes concerning the operation of the, Peltronic emission condenser, Notice – VACUUBRAND PC 3001 VARIOpro emission condenser Peltronic User Manual

Page 46

Advertising
background image

page 46 of 120

• Ensure that the

coolant outlet tubing at both con-

densers is always unobstructed and that it cannot get

blocked.

• Water pressure shock at the condensers may cause

breakage of the condenser coil!

Notes concerning the operation of the Peltronic emis-

sion condenser

Read the manual of the Peltronic emission condenser and

observe the instructions contained in that manual!

Assembling the Peltronic emission condenser:

Assemble the Peltronic emission condenser

with two attachment screws (Phillips screw-

driver size 2) to the pumping unit (a).

Assemble the hose connection between

pump outlet and condenser inlet (b).

Slip connecting tube onto hose connection

of the Peltronic emission condenser. Se-

cure with hose clip.

Connect the VACUU•BUS cable from the

CVC 3000 controller to the Peltronic emis-

sion condenser (c).

Notes concerning the operation of pumping units with

silencer

Attention: Dust-laden gases, deposits and condensed

solvent vapor can restrict air flow out the silencer. The re-

sultant back pressure can lead to damage of pump bear-

ings, diaphragms, and valves. Under those conditions, a

silencer must not be used.

NOTICE

(a)

(b)

(c)

NOTICE

Advertising