Adopt suitable measures to prevent the formation of

explosive or flammable mixtures, use inert gas for

venting if necessary.

Prevent the discharge of dangerous gases and vapors

to the surrounding atmosphere. If appropriate, connect

the exhaust line to a suitable treatment system.

Adopt suitable measures to prevent the release of

dangerous, toxic, explosive, corrosive, noxious or pol-

luting fluids, vapors and gases when disposing of con-

densates.

+

Cold surfaces can absorb a great amount of gas, which

may expand abruptly in case of warming. This may lead

to unmanageable overpressure in the system, creating

a risk of bursting!

+

Check the coolant level of the condenser at appropri-

ate intervals.

+

Attention: Coolant may overflow unexpectedly, e.g.,

in case of large amounts of gas.

+

Do not overfill the condenser. Lowest permissible cool-

ant temperature: -112°F (-80°C).

+

Maximum operating pressure: 16 psi (1.1 bar) abso-

lute.

+

Never block the gas outlet ((6) hose nozzle for tubing

I.D. 3/8” (10 mm)). The exhaust hose has always to be

unobstructed and without back pressure to enable an

unhindered discharge of gases and protect the pump

valves from damage.

• Prior to every use: Inspect the condenser for faults.

There must be no damage on the glass surfaces. Do

not use damaged components.

• Use only transport receptacles intended for coolants.

• Use safety glasses and protective gloves.

• Do not clamp the covers of condensers or coolant re-

ceptacles. Allow pressure equalization between cool-

ant and atmosphere.

• Comply with all applicable safety measures and re-

quirements when using cryogenic coolants.