Catchpots:

The catchpot at the inlet protects against drop-

lets and particles from entering the pump.
+

Enhances lifetimes of diaphragms and

valves.

+

Improves vacuum performance in applica-

tions with condensable vapors.

Both catchpots are coated with a protective

layer to protect against shattering in case of

breakage or implosion.

Assemble the catchpots at the inlet and at

the outlet using joint clips.

Exhaust waste vapor condenser

(PC3001 VARIO

pro

):

Assemble the hose nozzles for coolant inlet

(11) and coolant outlet (12) tubing and for

the gas outlet (6) at the exhaust waste va-

por condenser.

The

exhaust waste vapor condenser enables an effi-

cient condensation of the pumped vapors at the outlet.

+

No backflow of condensates.

+

Controlled recovery of condensates.

+

Close to 100% solvent recovery.

+

The isolation cover protects against glass splinters in

case of breakage, acts as thermal insulation to avoid

condensation of humidity and is intended to absorb

shocks.

Attach the tubing of the coolant circuit to the respective

hose nozzles (hose nozzles for tubing I.D. 1/4”-5/16”

(6-8 mm), see image) at the waste vapor condenser.

Check all hose connections prior to starting operation of

the cooling system.

Secure coolant hoses at the hose nozzles (e.g., with hose

clamps) to prevent their accidentally slipping off.

catchpot at

the inlet (10)

catchpot at the outlet (10)

11

12

6