VACUUBRAND PC 3001 VARIOpro emission condenser Peltronic User Manual

Page 41

Advertising
background image

page 41 of 120

- Condensate and deposits will affect the mea-

surement results. Clean the pressure trans-

ducer, if necessary. See section „Cleaning

the pressure transducer“, pg. 83.

Connecting components at the controller:

At the rear side of the controller are connec-

tions for e.g., an optional coolant valve and/or

an external pressure transducer and/or an ex-

ternal venting valve, as well as a venting con-

nection and the serial interface RS 232C.

- Pull the controller out of the housing. Be careful not to

disconnect the measuring connection (PTFE tube)!

- Insert the cable of the component to be connected un-

derneath the housing and plug into the controller. Do

not apply off-axis forces when assembling or removing

plug connections! Observe correct orientation of the

plug.

- Insert controller into the housing.

Separator at inlet and exhaust waste vapor condens-

er:

Assembling the hose nozzle with union nut:

Take the hose nozzle with attached compression fer-

rule and union nut out of the catchpot and put onto

inlet connection.

Tighten the union nut by hand until you can feel the

stop. Then tighten an additional 1/4 rotation with an

open-ended wrench (size 17mm) for final installation.

measurement

connection

at CVC 3000