Attention: At pressures above approximately 795 Torr

(1060 mbar) the pressure reading becomes incorrect

due to saturation of the pressure transducer. The dis-

play flashes. Release pressure immediately! Risk of

bursting!

+

Comply with applicable regulations when disposing of

chemicals. Take into consideration that chemicals may

be contaminated. Take adequate precautions to pro-

tect people from the effects of dangerous substances

(chemicals, thermal decomposition products of fluoro-

elastomers). Use appropriate protective clothing and

safety goggles.

+

Interruption of the pump (e.g., due to power failure),

failure of connected components or of parts of the sup-

ply, or change in parameters must not be allowed to

lead to dangerous conditions. In case of a diaphragm

failure or in case of a leak in the manifold, pumped

substances might be released into the environment or

into the pump housing or motor.

Comply with all notes regarding proper use of the

pumps, as well as operation and maintenance guid-

ance.

+

The residual

leak rate of the equipment might render

possible an exchange of gas, albeit extremely slight,

between the environment and the vacuum system.

Adopt suitable measures to prevent contamination of

the pumped substances or the environment.

PC 3001 TE VARIO

pro

:

+

Cold surfaces can absorb a great amount of gas, which

may expand abruptly in case of warming. This may lead

to unmanageable overpressure in the system, creating

a risk of bursting!

+

Check the coolant level of the condenser at appropri-

ate intervals.

+

Attention: Coolant may overflow unexpectedly, e.g.,

in case of large amounts of gas.