Cleaning and inspecting the pump heads

Remove catchpots (10) at inlet and outlet

(see „Use and operation“, pg. 38).

Detach the coupling of the hose connection

(M) below the housing cover (A).

Open the hose clip (N) with a slotted screw-

driver.

Pull the tubing off the hose connector.

Opening the hose clip:

Apply slotted screwdriver as shown and

turn.

N

M

A

PC 3001 VARIO

pro

/ PC 3001 VARIO

pro

IK

Disassemble the exhaust vapor condenser.

Loosen the union nut at the inlet of the ex-

haust vapor condenser.

Remove the 4 screws affixing the counter

holder (O) of the exhaust vapor condenser

with a Torx driver T10.

Remove the exhaust vapor condenser. In

doing so pull the PTFE hose out of the inlet

of the exhaust vapor condenser.

O