Replacing the diaphragm – VACUUBRAND PC 3001 VARIOpro emission condenser Peltronic User Manual

Page 105

Advertising
background image

page 105 of 120

Replacing the diaphragm

Remove the head cover (C) carefully from

the housing cover (A). Note position and

alignment of valves (B). Remove the valves.

+

Replace valves if necessary.

+

Use petroleum ether or other industrial sol-

vent to remove deposits. Do not inhale va-

pors.

+

Check diaphragms (E) for damage and re-

place if necessary.

Lift diaphragm carefully sidewise.

+

Never use a pointed or sharp-edged tool to

lift the diaphragm. If necessary, press down

the diaphragm clamping disc of the adja-

cent pump head to raise the diaphragm into

its uppermost turning point position.

Use the diaphragm key to grip the diaphragm

support disc (F) below the diaphragm.

Press down the diaphragm clamping disc to bring the diaphragm into

its lower turning point position. Unscrew diaphragm support disc (F)

with diaphragm (E) and diaphragm clamping disc (D).

Check for washers (G) between the diaphragm support disc (F) and

the connecting rod (H). Do not mix the washers from the different pump

heads, since these are set at the factory to ensure proper pump per-

formance. Make sure that the original number is reassembled at the

individual pump head.

+

Too few washers: The pump will not attain vacuum specification. Too

many washers: Diaphragm clamping disc will hit head cover, causing

noisy operation and possibly causing the pump to seize up.