Prevent the discharge of dangerous gases and vapors

to the surrounding atmosphere. If appropriate, connect

the exhaust line to a suitable treatment system.

+

Never block the gas outlet ((8) hose nozzle for tubing

I.D. 3/8” (10 mm)). The exhaust hose has always to be

unobstructed and without back pressure to enable an

unhindered discharge of gases and protect the pump

valves from damage.

Note: Install the hoses of the cooling system in a way

to avoid the flow / dripping of condensed water onto the

pumping unit (especially cables and electronic parts,

see also IP degree of protection, „Technical data“, pg.

27.

• Ensure that the

coolant outlet tubing is always unob-

structed and that it cannot get blocked.

• Maximum permissible coolant pressure at the exhaust

waste vapor condenser: 87 psi (6 bar) absolute. Outlet

flow must always be unhindered.

• Comply with the maximum permissible coolant pres-

sures of additional components in the coolant circuit

(e.g., coolant valve).

• Avoid overpressure in the coolant circuit (e.g., caused

by blocked or squeezed coolant hoses).

• Only install the optional coolant valve in the supply line

of the exhaust waste vapor condenser.

Notes regarding the (dry) ice condenser

(PC 3001 TE VARIO

pro

)

The

exhaust waste vapor condenser en-

ables an efficient condensation of the pumped

vapors at the outlet.
+

Close to 100% solvent recovery.

+

The isolation cover protects against glass

splinters in case of breaking, acts as ther-

mal insulation to avoid the condensation of

humidity and is intended to absorb shocks.

outlet (gas!)

hose nozzle