Use the equipment

only as intended, that is, for conden-

sation of vapors at the pump outlet or inlet.

Suitable coolants: e.g., dry ice or ethanol-dry ice mixture

or water-ice mixture.

Remove cap. Fill coolant into the coolant

chamber. Put cap back in place. Do not

clamp the cap.

+

Check coolant level at appropriate intervals.

Check level of condensate at appropriate

intervals and drain condensate if necessary.

+

Wait until the condensate has liquefied, if

necessary.

+

Separate the inner part of the condenser

from the vacuum installation to drain the

coolant.

Remove the cap on top of the condenser.

Turn the inner part of the condenser and lift

to pull it out. Comply with regulations when

disposing of chemicals.

+

When reassembling check for the correct

position of the seal between inner and outer

parts of the condenser.

Notes concerning the operation of the inlet condenser

IK

+

Attention: The inlet condenser has no overpressure

safety relief device.

Note: Install the hoses of the cooling system in a way

to avoid the flow / dripping of condensed water onto the

pumping unit (especially cables and electronic parts),

or use thermally insulated hoses to prevent condensa-

tion.