Gas inlet temperatures

Wetted parts

Components

Wetted materials

Pump

Housing cover

PTFE

Head cover

ETFE carbon fiber reinforced

Diaphragm clamping disc

ETFE carbon fiber reinforced

Diaphragm

PTFE

Valve

FFKM

Pumping unit
Inlet pumping unit

PPS (PC 3001 VARIO

pro

IK: PP)

Outlet pumping unit

PET (PC 3001 without emission con-

denser: PTFE carbon reinforced)

Outlet pumping unit (PC 3001 VARIO

pro

emission condenser Peltronic)

ETFE / ECTFE

Tubing

PTFE

Screw-in fittings

ETFE / ECTFE

O-ring at the catchpot

Fluoroelastomer

Overpressure safety relief device

at exhaust waste vapor condenser

Silicone rubber / PTFE film

Distribution head (inlet)

PPS glass fiber reinforced /

PP (blind plug)

Waste vapor condenser IK / EK / TE,

catchpot

Borosilicate glass

Emission condenser Peltronic

ETFE / ECTFE / PP / PA

Silencer

PBT / PVF / rubber

CVC 3000

Sensor

Aluminum oxide ceramic

Sensor housing

PPS / glass fiber

Sensor seal

Chemically resistant fluoroelastomer

Venting valve seal

FPM

We reserve the right for technical modification without prior notice!

* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)

Operating condition

Inlet pressure

Permitted range of gas

temperatures at inlet

Continuous operation

> 75 Torr (100 mbar)

(high gas load)

50 °F to 104 °F

(+10°C to +40°C)

Continuous operation

< 75 Torr (100 mbar)

(low gas load)

32 °F to 140 °F*

(0°C to +60°C*)

Short-time

(< 5 minutes)

< 75 Torr (100 mbar)

(low gas load)

14 °F to 176 °F*

(-10°C to +80°C*)

Advertising