Service only one side of the pump at a time to avoid the mixing of

parts.

PC 3001 VARIO

pro

IK

Disassemble the inlet condenser.

Loosen the union nut at the inlet of the

pump.

Remove the 4 screws affixing the counter

holder of the exhaust vapor condenser with

a Torx driver T10.

Remove the inlet condenser. In doing so

pull the PTFE hose out of the inlet of the

inlet condenser.

PC 3001 VARIO

pro

TE

Disassemble the emission condenser TE.

Loosen the union nut at the inlet of the emis-

sion condenser TE.

Remove the 2 screws affixing the mounting

plate of the emission condenser TE with a

Phillips screw driver size 2.

Remove the emission condenser TE. In do-

ing so pull the PTFE hose out of the inlet of

the emission condenser TE.

PC 3001 VARIO

pro

EK Peltronic

Disassemble the Peltronic emission condens-

er. Detach the coupling of the hose connection

at the Peltronic emission condenser.

Open the hose clip with a slotted screw-

driver.

Pull the tubing off the hose connector.

Remove the 2 screws affixing the mounting

plate of the Peltronic emission condenser

with a Phillips screw driver size 2.

Remove the Peltronic emission condenser.

In doing so pull the PTFE hose out of the

inlet of the Peltronic emission condenser.