Assembling the housing cover, At the side of the emission condenser, Disassembling the housing cover – VACUUBRAND PC 3001 VARIOpro emission condenser Peltronic User Manual

Page 108: At the side of the on/off switch

Advertising
background image

page 108 of 120

Assembling the housing cover at the side of the emis-

sion condenser

Put on housing cover (A).

+

Move housing cover slightly to ensure that the head covers are cor-

rectly positioned.

Screw in the Allen head screws at the head

cover in a diagonal pattern with a 4 mm

wide Allen key, loosely at first, to align, then

tighten.

+

Maximum torque: 4.4 ft

.

lb

f

(6 Nm).

Disassembling the housing cover at the side of the ON/

OFF switch

Lay the pumping unit on its side with the

pump heads to be maintained at the top.

Support the pumping unit appropriately.

Disassemble the housing cover (A) to check

the valves (B).

Unscrew six Allen screws with a 4mm wide

Allen key. Remove the housing cover with

housing cover insert (A) together with head

covers (C) and valves (B).

+

Never use a pointed or sharp-edged tool to remove parts (e.g., screw-

driver). We recommend using a rubber mallet or compressed air (to be

blown carefully into port).

Remove the head cover (C) carefully from

the housing cover (A). Note position and

alignment of valves (B). Remove the valves.

+

Replace valves if necessary.

+

Use petroleum ether or other industrial sol-

vent to remove deposits. Do not inhale va-

pors.

Advertising
This manual is related to the following products: