exhaust waste vapor condenser (9) regularly; replace

if necessary. Check especially for deterioration, co-

alescence and cracks.

• Ensure that the

coolant outlet hose is always free

and that it cannot get blocked.

• Maximum permissible coolant pressure at the exhaust

waste vapor condenser: 87 psi (6 bar) absolute

• Comply with the maximum permissible coolant pres-

sures of additional components in the coolant circuit

(e.g., coolant valve).

• We strongly recommend installing an optional coolant

valve (see „Accessories“, pg. 94)

in the supply line

of the exhaust vapor condenser to save water and re-

duce the risk of water spill.

• Avoid overpressure in the coolant circuit (e.g., caused

by blocked or kinked coolant hoses).

• Avoid any clogging of the exhaust vapor condenser

caused by deposits or frozen solvents. Clean if neces-

sary or use higher coolant temperatures.

In case of

condensation: Check the liquid level in both

catchpots (10) during operation. Check the liquid lev-

el regularly. Do not allow the catchpots to overfill. Drain

catchpots in time to avoid overflow. Install a level sensor

(see „Accessories“, pg. 94) for monitoring, if necessary.

The maximum liquid level is at approximately 80% of the

total filling level to avoid problems when removing the

catchpots.

Permissible range of coolant temperature at the exhaust

waste vapor condenser:

5 °F to 68 °F (-15°C to +20°C)

Check hose connections prior to starting operation of the

cooling system.

Check coolant hoses regularly during operation.

Removing the catchpots:

Stop process.

NOTICE