5.1 Installation site

Operate the freeze-dryer only in closed and dry rooms.
• The table must be stable and have a solid, even tabletop.
• Ensure sufficient ventilation. Do not place any paper, cloth, or similar

material behind or under the unit, since otherwise the air circulation will
be impaired.

• Keep a safety distance of at least 30 cm from the wall so that the vents

in the unit remain fully effective.

• The ambient temperature must be in the range of +10°C to +25°C.
• Do not subject the freeze-dryer to thermal stress, e.g. by positioning it

near heat generators.

• Avoid direct sunlight (UV radiation).

Pos: 92 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 93 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /05 0 Aufstell ung und Anschlus s/05 0-0 020 Ene rgiev erso rgu ng- -- --- --- -- --- -- --- --- -- --- -- --- -- --- -- @ 6\ mod _129 913 713 915 6_6 8.doc x @ 3 499 8 @ 2 @ 1

5.2 Power supply

Pos: 94 / 200 Christ /36 1 GT -BA La bor -Pilot (PROJEKTE )/Beta 1- 8_2 -8 LSCpl us/0 50 Aufs tellun g un d Ansc hluss/0 50- 002 0-0 010 Anschl ussa rt Beta @ 1 6\m od_1 359 106 620 131 _68. docx @ 80 782 @ 3 @ 1

5.2.1 Connection

DANGER

The operating voltage on the name plate must correspond to the local
supply voltage


CHRIST freeze-dryers are units of safety class I. The freeze-dryers of the
type Beta 1-8 LSC

plus

and Beta 2-8 LSC

plus

have a three-wire power cord

with an IEC C13 connector (see chapter 10 - "Technical data").
An equipotential bonding screw is located on the back below the mains
power input (see chapter 2.1.1 - "Functional and operating elements").
This equipotential bonding screw can be used to perform an earth
conductor check.

Pos: 95 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 96 / 200 Christ /36 0 GT -BA La bor -Pilot (S TANDARDMODULE) /05 0 Aufstell ung und Anschlus s/05 0-0 020 -00 20 Sic her unge n b auseits @ 6 \mo d_1 299 1371 398 43_ 68.d ocx @ 350 14 @ 3 @ 1

5.2.2 Customer-provided fuses

Typically, the freeze-dryer must be protected with 16 Amp G fuses that are
to be provided by the customer.