Beta 1-8 LSC&
Beta 2-8 LSC&

5 Set-up and connection

32

Version 04/2013, Rev. 1.1 of 16/12/2013 • sb

Translation of the original operating manual

Pos: 98 / 200 Christ /36 1 GT -BA La bor -Pilot (PROJEKTE )/Beta 1- 8_2 -8 LSCpl us/0 50 Aufs tellun g un d Ansc hluss/0 50- 003 0 Belü ftun gsven til Beta @ 16\ mo d_1 3591 066 209 93_ 68.d ocx @ 807 98 @ 2 @ 1

5.3 Aeration valve

The aeration valve is located on top of the left side of the unit (see chapter

2.1.1 - "Functional and operating elements").
After the end of a freeze-drying process, the unit will be aerated via the
aeration valve.

NOTE

The ice condenser chamber can be flooded with nitrogen via the hose
nozzle of the aeration valve.

WARNING

For the aeration with gases under pressure, the client must install a safety
device for pressure relief.

Pos: 99 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 100 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 101 /20 0 Ch rist/3 61 G T-BA L abo r-Pilot (PROJEKTE)/Bet a 1 -8_ 2-8 LSCplus/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 40 Me diena blauf ventil Be ta @ 16\ mod _13 591 066 212 57_6 8.d ocx @ 808 06 @ 2 @ 1

5.4 Media drain valve

The media drain valve is located at the bottom of the left side of the unit
(see chapter 2.1.1 - "Functional and operating elements").
It is used to drain off the condensate and the defrosting water.
• Connect the drain hose (included in the scope of supply) to the hose

connector.

• Place a collecting vessel under the unit.
The hose must be laid with a continuous slope and the end of the hose
must always be above the liquid level in the collecting vessel. This prevents
water and dirt residues from being sucked into the ice condenser chamber
if there is negative pressure when the media drain valve is opened.

Pos: 102 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 103 /20 0 Ch rist/3 61 G T-BA L abo r-Pilot (PROJEKTE)/Bet a 1 -8_ 2-8 LSCplus/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 50 Va kuu mmes sson de Bet a @ 16\m od_ 135 910 662 034 8_68 .docx @ 8 079 0 @ 2 @ 1

5.5 Vacuum sensor

NOTE

Please refer to the separate operating manual of the vacuum sensor!


In order to protect the vacuum sensor against transport damage, it comes
supplied in its original packaging. Prior to commissioning the freeze-dryer,
the sensor must be installed.