7 Malfunctions and error correction

Version 04/2013, Rev. 1.1 of 16/12/2013 • sb

77

Translation of the original operating manual

Pos: 170 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010 -00 20- 0020 Belüf tun gsven til, Me dien abla ufven til @ 6\ mo d_12 991 375 161 22_ 68.d ocx @ 351 52 @ 4 @ 1

7.1.2.2 Aeration valve, media drain valve

A malfunction of the aeration valve or the media drain valve may have
several causes. One potential source are contaminants such as product
residues within the valve.
• Switch the freeze-dryer off and disconnect the mains plug.
• Clean the valve (see chapter 8.1.1.3 - "Aeration valve, media drain

valve").

• Put the freeze-dryer into operation again.
If there is still a leakage, the freeze-dryer must be checked by qualified
specialist personnel (see chapter 7.3 - "Service contact").

Pos: 171 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 172 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010 -00 20- 0021 Dr uckste uerv entil @ 15\ mo d_1 3578 991 634 37_ 68.d ocx @ 790 39 @ 4 @ 1

7.1.2.3 Pressure control valve

A malfunction of the pressure control valve may have several causes.

NOTE

The inspection of the valve must be carried out by qualified specialist

personnel (see chapter 7.3 - "Service contact").

Pos: 173 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 174 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010 -00 20- 0030 G ummiv entile @ 6\ mo d_12 991 375 164 66_ 68.d ocx @ 351 60 @ 4 @ 1

7.1.2.4 Rubber valves

In order to identify a leaking rubber valve, the valves must be checked
individually:
• Remove the rubber valve and seal the connection at the drying

chamber with a rubber stopper.

• Check the tightness under vacuum until the leaking valve has been

localised.

• Clean the valve or replace it if necessary.

Pos: 175 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 176 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010 -00 20- 0040 Vaku um messs ond e @ 6 \mo d_1 299 137 5169 66_ 68. docx @ 351 68 @ 4 @ 1

7.1.2.5 Vacuum sensor

Vacuum sensors have a limited service life and can be ordered as spare
parts (e.g. VSP62/63, order no. 125468).
Functional test:
• Connect the vacuum sensor directly to the suction side connector of the

vacuum pump.

If a final pressure of at least 0.011 mbar is reached (with a vacuum pump
that has reached its operating temperature), the sensor and the vacuum
pump are OK.