6 Operation

42

Version 04/2013, Rev. 1.1 of 16/12/2013 • sb

Translation of the original operating manual

Option: USB process recording (see chapter 6.6 - "Optional extensions")
• Select the "Process recording" function in the dialog box "Tools".
• Select the input fields ("Batch data"). A keyboard for the data input will

be displayed.

• If necessary, select the “Options” tab, choose between manual or

automatic recording, and define a recording interval.

• Press the "return" button in order to close the dialog box.
The process recording will now run in the background.

Fig. 25: Dialog box “USB process recording”


Option: Pressure increase test (only with the double-chamber method
(process B)
)
The pressure increase test can only be performed when the freeze-dryer
is equipped with an intermediate valve. The performance is possible in the
manual mode as well as in the program mode (see chapter 6.5.3.1 -
"Creating a program").
• Select “Pressure increase test” in the “Tools” dialog box.
• Enter the set values for the duration and maximum pressure increase

with the aid of the buttons.

• Start of the pressure increase test. The test time will be displayed. After

the end of the test, a status message (pressure increase was success-
ful or failed) will be displayed together with the measured values: