Martin Christ Beta 2-8 LSCplus User Manual

Page 89

Advertising
background image

Beta 1-8 LSC&

Beta 2-8 LSC&

10 Technical data

Version 04/2013, Rev. 1.1 of 16/12/2013 • sb

89

Translation of the original operating manual

Connection requirements
(without vacuum pump and

accessories)

Beta 1-8 LSC

plus

Beta 2-8 LSC

plus

Electrical connection:

1 x 230 V / 50 Hz
(other supply data on request)

1 x 230 V / 50 Hz
(other supply data on request)

Protection class:

I

I

IP protection category according to

DIN 60529:

11

11

Nominal power:

0.92 kVA

1.1 kVA

Nominal current:

4.0 A

5.0 A

Mains fuse:

10 A

12 A

Pressure control valve connection: 230 V, 50/60 Hz, 20 VA max.

230 V, 50/60 Hz, 20 VA max.

Vacuum pump connection:

230 V, 50/60 Hz, 2 A max.

230 V, 50/60 Hz, 2 A max.

Filling quantity:

Beta 1-8 LSC

plus

Beta 2-8 LSC

plus

Refrigerant

See name plate

See name plate

Physical data (without vacuum
pump and accessories):

Beta 1-8 LSC

plus

Beta 2-8 LSC

plus

Dimensions
- height:
- width:
- depth:


415 mm
780 mm
540 mm
(incl. vacuum flange connection)


415 mm
780 mm
540 mm
(incl. vacuum flange connection)

Weight:

approx. 63 kg

approx. 78 kg

Noise level according to DIN
45635:


54 dB(A)


54 dB(A)

EMC according to EN 55011:

Class B

Class B

Heat emission:

0.6 kW min.
1.1 kW max.

1.1 kW min
1.6 kW max.

Equipment connections:

Beta 1-8 LSC

plus

Beta 2-8 LSC

plus

Vacuum connection:

Small flange connection DN25KF
(ISO 28403, DIN 2861)

Small flange connection DN25KF
(ISO 28403, DIN 2861)

Media drain valve:

Hose nozzle DN10
(outside diameter 12 mm)

Hose nozzle DN10
(outside diameter 12 mm)

Aeration valve:

Hose nozzle DN4

(outside diameter 8 mm max.)

Hose nozzle DN4

(outside diameter 8 mm max.)

Vacuum sensor:

SUB D-9
VSP 62

SUB D-9
VSP 62

Data interface (LAN):

RJ 45

RJ 45

Special equipment: water
cooling system

Beta 1-8 LSC

plus

Beta 2-8 LSC

plus

Advertising