6 Operation

Version 04/2013, Rev. 1.1 of 16/12/2013 • sb

73

Translation of the original operating manual

Lyo Control LC

plus

measuring system

The Lyo Control LC

plus

measuring system can be used to determine the

crystallisation state of the product. In the liquid state, the electrical
resistance is very low. During the freezing process, the resistance
increases. The LyoRx control sensor measures the electrical resistance.

LyoLog

plus

data logging software

LyoLog

plus

is a data logging software program by Martin Christ

Gefriertrocknungsanlagen GmbH that is specifically adapted to the
requirements of freeze-drying processes. Apart from the graphical
representation of the measurement data of currently running processes, it
also enables the data export for additional evaluation.

LPC

plus

SCADA system

The Christ LPC

plus

system consists of the Supervisory Control And Data

Acquision (SCADA) software program by Martin Christ Gefriertrocknungs-
anlagen GmbH and a dedicated PC. The system is connected to the
LSC

plus

controller of the freeze-dryer via Ethernet LAN and provides the

operation of all of the freeze-drying functions as well as process recording
(measurement data and process events), process documentation and data
backup. Furthermore, it enables the comfortable administration of freeze-
drying programs/recipes and users.

Pos: 154 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 155 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 156 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 60 Bet rieb/ 060 -00 70 A usschalt en Alp ha - Ga mma ++ ++ ++ +++ ++ ++ ++ +++ ++ ++ +++ ++ @ 6\ mod _13 028 463 824 33_6 8.d ocx @ 399 88 @ 2 @ 1

6.7 Switching the freeze-dryer OFF

The freeze-dryer must be in the standby status.
• Switch the freeze-dryer off by pressing the mains power switch.