6 Operation

64

Version 04/2013, Rev. 1.1 of 16/12/2013 • sb

Translation of the original operating manual

Vacuum pump warm-up
Apart from the execution of the various sections, the PGM

plus

programmer

module also controls other tasks that need to be performed during the
process run, e.g. the automatic activation of the vacuum pump.
If the pressure control valve is closed, the PGM

plus

programmer module

shifts the warm-up phase of the vacuum pump to the freezing phase so that
it is before the first drying section. During the warm-up of the vacuum
pump, neither the freezing phase nor the pressure inside the drying
chamber will be affected.
The duration of the warm-up phase can be specified separately for every
program (see chapter 6.5.3.1 - "Creating a program").

Fig. 59: The warm-up phase of the vacuum pump is shifted to the freezing phase prior to the

first drying section

Pos: 141 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 142 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 60 Bet rieb/ LSCplus/ 060 -00 50- 003 2-0 010 Prog ra mm erstell en L abo r @ 6\m od_ 129 9142 755 014 _68. docx @ 36 719 @ 4 @ 1

6.5.3.1 Creating a program

To create a new program:
• Press the button "New program" in the main window "Program". The

dialog box that is displayed offers various different program templates.
The scroll bar on the right-hand side can be used to scroll through the
list.

1 Button for creating a

blank program

2 Scroll bar for

selecting a program

template

Fig. 60: Dialog box "Select program template"

T/°C

t/h

100

10

1

0.1

0.01

0.001

+30

+20

+10

0

-10

-20

Pressure/mbar

Section

“freezing”

First

section

“drying”

Warm-up phase

of the vacuum pump

Last

section

“freezing”