Martin Christ Beta 2-8 LSCplus User Manual

Page 68

Advertising
background image

Beta 1-8 LSC&
Beta 2-8 LSC&

6 Operation

68

Version 04/2013, Rev. 1.1 of 16/12/2013 • sb

Translation of the original operating manual

Option: Comparative pressure measurement (see chapter 6.6 - "Optional
extensions
")

NOTE

The indication of the measurement value "dp Comp" can be configured in

chapter 6.5.1.1 - "Main window "Manual"" under "Actual values of the
current process" in the dialog box "Select actual value".


The comparative pressure measurement can be activated or deactivated by
pressing the button in the dialog box.

Progress condition:

If the actual value is below the specified limit, the system will switch
from the last drying section of the main drying phase to the final drying
phase. If several conditions for switching to the next section are

activated (chapter 6.5.3 - "Option: freeze-drying with the PGMplus
programmer
module", section "Conditions for switching to the next
section"), the system will not switch to the next section until all of the
conditions are fulfilled.

Disabled:

There will be no comparative pressure measurement during the main
drying phase. There will be no evaluation.

Fig. 64: Dialog box "Comparative pressure measurement"


Program templates (2)
The PGM

plus

programmer module offers 14 different program templates that

include recipes for all kinds of freeze-drying applications. They are
suggestions for orientation purposes and must be adapted to the specific
area of application of the freeze-dryer.
Following the selection of a program template, a window containing
information concerning the drying process will be displayed.

Advertising
This manual is related to the following products: