Martin Christ Beta 2-8 LSCplus User Manual

Page 44

Advertising
background image

Beta 1-8 LSC&
Beta 2-8 LSC&

6 Operation

44

Version 04/2013, Rev. 1.1 of 16/12/2013 • sb

Translation of the original operating manual

Malfunctions are divided into three categories:
• Red:

error messages

• Orange: process messages
• Yellow: general messages
The representation of the message provides information on its current
status. A double frame around a message means that the error has not
been eliminated yet. The colour of the button "quit" changes from blue to
grey once the message has been acknowledged.

Fig. 28: Representation of an error message


The advantage of this system is that malfunctions that occurred during the
night can be discovered the next day even if the cause of the malfunction
has already been eliminated.
The dialog box cannot be quit until all of the messages have been
acknowledged.
If a message has been acknowledged although the malfunction has not
been eliminated, the button "Process and equipment messages" will be
displayed in the respective colour of the malfunction in the main window.


1

The malfunction has
been eliminated, but
the message has not
been acknowledged

yet

2

The malfunction has
not be eliminated and
the message has not
been acknowledged

yet

3

The malfunction has
not been eliminated
yet, but the message
has been

acknowledged

Fig. 29: Dialog box "Process and equipment messages"


Details
Touching the message calls up details concerning the error message:
• Cause of the message
• Effects of the message
• Measure to eliminate the error

Reference designator

Use the arrow keys to call up the individual windows.
The error message text is always followed by an error code.

Advertising
This manual is related to the following products: