2 remote sense connections, Remote sense connections – AMETEK SLH Series RevC User Manual

Page 56

Advertising
background image

Applications

Sorensen SLH-Series DC High Power Electronic Load

4-2

M540074-01

Rev C

4.2

REMOTE SENSE CONNECTIONS

Figure 4-2 illustrates a typical set up with the electronic load connected for remote sense
operation. The remote V-sense BNC cable of the electronic load are connected to the output of
the power supply. Remote sensing compensates for the voltage drop in applications that
require long lead lengths.
The 4½-digit volt Meter of SLH-series DC electronic load measures the voltage of V-sense BNC
input Terminal automatically, so the high accuracy 5-character voltage Meter can measure the
specific points voltage of the power supply's output voltage.
Load leads should be bundled or tie wrapped together to minimize inductance.

Figure 4-2 Remote Voltage Sense Connections

Advertising