2 query commands – AMETEK SLH Series RevC User Manual

Page 75

Advertising
background image

Sorensen SLH-Series DC High Power Electronic Load

Remote Programming

SHORT ON/OFF

Purpose:

Short the DC input of Electronic load module.

Command Syntax:
SLH DC, SLD Modules: [GLOB:]SHOR{SP}{0|1|OFF|ON}{NL}

Description:
This command applies the short across the input of the Electronic load. Executing SHOR does not
affect any programmed settings and the Electronic load will return to those settings when the short
is removed.

Example:
SHOR ON ; set load input to short state.
SHOR OFF ; set load input short open state.

SENSE ON/OFF

Purpose:

Set the voltage sense AUTO/ON or OFF of Electronic load.

Command Syntax:
SENS{SP}{0|1|OFF|ON}{NL}

Description:
Set the Vsense inputs Off or On. If Vsense is set to Off, voltage measurement is taken at the load
input terminals. On SLH-seris DC loads, Vsense is auto/off or on. If it is set to auto/off, the module
will measure the Vsense input if the voltage is greater than 1V (on 60V models).

Example:
SENS ON set all load V-sense input to sense AUTO.
SENS OFF set load V-sense input to sense OFF state.

DYNAMIC ON/OFF

Purpose:
Set DYNamic ON or OFF command.

Command Syntax:
DYN{SP}{0|1|OFF|ON}{NL}

Description:
DYN OFF ; set the load to static load mode.
DYN 1 ; set load to dynamic load mode.

5.3.2 Query

Commands

CURRENT Level

Purpose:
The Constant Current mode's load current level query command.

Command Syntax:
CC:{HIGH|LOW}{?}{;|NL}

Description:

M540074-01

Rev C

5-13

Advertising