4.4

CONSTANT VOLTAGE (CV) MODE APPLICATION

Major applications:

Current source testing, power supply current limit characteristic testing

Current Source Testing

Battery chargers are among the most common current sources; they are used to inputs current
to a re-chargeable battery. The CV mode of electronic load can be used to simulate the
terminal voltage of re-chargeable battery, to test the current being generated by the battery
charger.

For example, CV mode can be used to first set voltage and read the charge current from the
current meter on the load; then set CV voltage to 5V (example), then read the charge current
again, this method is used to test the load regulation of a current source.

Current Limit Testing

A lab power supply could use the fold-back current limit test for a switching power supply, to
find the constant current limit curve. CV mode is effective for such testing.

For example, the user can set CV voltage, record the output current, then run a series of
voltage and current to get an output current limit curve (Figure 4-5) of a power supply.

Figure 4-5 Constant Voltage Mode Application