Section B - Linear Glass Scale Installation

Mounting Surface Preparation

B1.0

Inspect the mounting surface for any machining
irregularities. MicroE Systems recommends a surface
finish of better than 3.2 micrometers Ra.

B1.1

MicroE Systems recommends a mounting surface
flatness of 0.0001 inch/inch.

B1.2

Thoroughly clean the scale mounting surface and
reference edge using a cotton swab or lint-free cloth
dampened with isopropyl alcohol or acetone.

Remove all dust and particles.