Vcm, vcm-x & vcb-x controllers, Vcb-x controllers, Accessing and entering vcb-x setpoints – Orion System OE392-10 System Manager TS User Manual

Page 35

Advertising
background image

SMTS Technical Guide

VCM, VCM-X & VCB-X Controllers

35

VCB-X Controllers

Accessing and Entering VCB-X

Setpoints

While in the VCB-X Status Screen (see Figure 35), touch the

< Set-

points>

button found on the bottom menu bar. The VCB-X Controller’s

Temperature Setpoints Screen will appear. See Figure 36.

Level 1 and Level 2 users can change occupied space temperature set-
points, but only Level 3 users can change all setpoints.

Accessing and Entering VCB-X Setpoints

Figure 35: VCB-X Controller Status Screen

Figure 36: VCB-X Controller Temperature Setpoints
Screen

Individual setpoint and confi guration buttons are located at the bottom
of the Setpoint Screens. See Figure 37. Simply touch a specifi c button
to access that category.

Within each Setpoint Screen, touch the

< Status>

button to return to the

Status Screen or touch

< Home>

to return to the Main Screen.

Use the

<+>

and

<->

buttons to scroll through the setpoints and confi gu-

rations. Simply touch the blue highlighted box to change the setpoint.
Each setpoint data entry screen will provide a defi nition of the setpoint
and specifi c instructions for entering the setpoint and will include the
setpoint range as in the example below, Figure 38.

Touch

<OK>

to have the system accept the new value. If you enter a

setpoint that is not in the valid range, the setpoint will remain as is and
will not change.

Each setpoint data entry screen is self-explanatory; however, each
setpoint and confi guration is explained in the next section, Confi guring
VCB-X Setpoints. To easily reference a particular setpoint or confi gura-
tion, refer to Index B.

Figure 37: VCB-X Setpoint Buttons

Temps

Static

Staging

Misc

Relays

Config

Figure 38: VCB-X Cooling Mode Setpoint Data
Entry Screen

Advertising