10

Verifying that Unit is Calibrated

Selecting CALIB sets up the mode to measure reflectance. This also allows you to
check the calibration of the unit by placing the wand on a template (white or black
field) and verifying that the reflectance value matches the stated values.

Verify that reflectance values are within 5% for white and 2% for black. If the values
are not within these tolerances you need to calibrate the unit; otherwise, press Enter
and Select at the same time to navigate back to the Main Menu.

Calibrating the Unit

After selecting Calib from the main menu, the reflectance menu will display.

Inspector 4000

Reflectance XX%

From the

reflectance

menu press Enter, the Calibrate menu

will

display:

Calibrate

When the “Calibrate” menu is displayed, proceed with the following calibration

steps

:

Place Optics on

White/Enter Key

Place the

wand

on the white section of the calibration plaque with the shoe

guide

flat

against the surface. Press Enter and hold auto-optic scanner head still until you hear a

“chirp” sound.