RJS Inspector D4000 Auto Optic (FIRMWARE version A.05 and Earlier) User Manual

Page 18

Advertising
background image

Operator

s Guide

Inspector 4000

002-5564

RJS, Minneapolis, MN

11

Place Optics on

Black/Enter Key

Place the

wand

on the black section of the calibration plaque with the shoe

guide

flat

against the surface. Press Enter and hold auto-optic scanner head still until you hear a

“chirp” sound.

When both light and dark have been calibrated the following screen will display:

Calibrated

Reflectance XX%

Verify that the unit is calibrated by again placing the scan head on the light or dark area of
the plaque.

The reflectance % value will display. Reflectance values should be within 5% (of the
known value of the plaque) for white and 2% (of the known value of the plaque) for black.

If the “Unable to Calibrate” screen displays, repeat the calibration.

Unable to

Calibrate

Note: Power drain is greatest in the calibration mode. The automatic shut-off feature is disabled in

calibration mode. Once the unit is calibrated, the calibration data will remain in memory;
however, you should calibrate the unit at the beginning of each day or each shift for the proper
PCS readings.

5.2 Scan

After finishing calibration

press

Select and Enter at the same time; the “Ready Menu”

will

display

: