Seaward 300 PrimeTest User Manual

Page 14

Advertising
background image

PrimeTest 300

Operating Instructions

- 13 -

2.0

Safety Notes

This PrimeTest 300 has been built and tested in
accordance with:

IEC 61010-1: 2001.
IEC 61557 part 1, 2, 4 and 10.

To ensure safe operation of the unit, all notes and
warnings in these instructions must be observed at all
times.

This instrument may be used on electrical
circuits

which

comply

to

over

voltage

category II.

Attention! During Insulation Resistance tests
high voltage levels occur within the PrimeTest
300.

The tester and all associated cables and leads
must be checked for signs of damage before

equipment is operated.

Where safe operation of the tester is no longer possible
it should be immediately shut down and secured to
prevent accidental operation.

It must be assumed that safe operation is no longer
possible:
- if the instrument or leads show visible signs of damage
or
- the instrument does not function or
-

after

long

periods

of

storage

under

adverse

environmental conditions.

Never attempt to charge non-rechargeable
cells in the PrimeTest 300.

Always check that the correct NiMH batteries
are fitted before connecting the charger.

Only use the Seaward PT- 01 charger to
charge the batteries in the PrimeTest 300.