- 83 -

15.0 Environmental Conditions

15.1 The PrimeTest 300 has been designed to perform

tests and measurements in a dry environment.

15.2

Maximum

barometric

elevation

for

making

measurements is 2000M.

15.3 Overvoltage category IEC 60664/IEC 61010, 300V

Category II.

15.4 Pollution degree 2 according to IEC 61010-1.

15.5 Protective system IP40 according to IEC 60529.

15.6 Electromagnetic compatibility (EMC). Interference

immunity and emitted interference conforming to
IEC 61326-1.

15.7 Operating temperature range of 0°C to 40°C,

without moisture condensation.

15.8 The

PrimeTest

300

can

be

stored

at

any

temperature in the range -25°C to +65°C (relative
humidity up to 90%). The batteries should be
taken out of the instrument for storage.