Seaward 300 PrimeTest User Manual

Page 82

Advertising
background image

PrimeTest 300

Operating Instructions

- 81 -

14.4

Class II Leakage Current

Test Voltage

Mains supply 230V +10% / -15%

Max Load Current

10A

Input Impedance

2KΩ As per IEC 61010-1 A.1

Test Duration

2, 5, 10, 20, 30 seconds

Pass/Fail Limit

User Defined

Display Range

0.00 – 3.50mA

Measuring Range

0.10 – 3.50mA

Resolution

0.01mA

Accuracy

± 5% ±5 digits

Attention: Avoid prolonged, repeated use at
full load and excessive test duration as this

may reduce the life of the unit.

14.5 Class

I Leakage Current

Measurement Method

Differential

Test Voltage

Mains supply 230V +10% / -15%

Max Load Current

10A

Test Duration

2, 5, 10, 20, 30 seconds

Pass/Fail Limit

User Defined

Display Range

0.00 – 19.99mA

Measuring Range

0.25 – 19.99mA

Resolution

0.01mA

Accuracy

± 5% ±5 digits

Attention: Avoid prolonged, repeated use at
full load and excessive test duration as this

may reduce the life of the unit.

14.6 Earth

Loop Resistance

Test Voltage

Mains supply 230V +10% / -15%

Test Current

Non-trip for 30mA RCD

Test Duration

Approx 10 seconds

Pass/Fail Limit

User Defined

Display Ranges

0.0 – 199Ω

Measuring Range

1.0 – 199Ω

Resolution

0.1Ω, 0.0 – 19.9Ω
1Ω,

20 – 199Ω

Accuracy

± 10% ±

1.2

1.2Ω