Pump operation, Carbon calculation – Super Systems MGA6010 User Manual

Page 11

Advertising
background image

MGA 6010 Operations Manual

Super Systems Inc.

Page 11 of 54

Pressing the Return button at the bottom right of the operator interface on any screen will take the

display to the default. It may be necessary to press the Return button multiple times.

Pump Operation

Initially, the pump will be off. The pump should remain off while sampling an endothermic generator or

any other atmosphere under positive pressure. The pump should also remain off during calibration. For

proper operation, the flow of gas through the sensors should be between 1.0 and 2.0 SCFH. If the flow

meter on the right of the screen or on the inside of the lid does not indicate sufficient flow, turn the pump

on. When accessed from the main screen, the pump has two possible modes: Automatic and Off.

Carbon Calculation

The MGA 6010 determines the percent carbon in the sample gas using measured amounts of CO, CO

2

, and

CH

4

along with the Furnace Temperature. The Furnace Temperature is either entered by the user or

obtained automatically from the Furnace Temperature Controller via RS-485 communications.

Additionally, the carbon percentage measured by the MGA 6010 can be used as a comparison to the

carbon percentage measured by a furnace’s oxygen probe. This is accomplished either by manually

entering the Probe Temperature, Probe Millivolts, and the Probe CO Factor into the MGA 6010 or by

obtaining the information automatically via RS-485 communications to the Furnace Carbon Controller.

Providing the probe information allows the MGA 6010 to suggest an adjustment for the probe CO Factor

(or Process Factor) in order to keep the oxygen probe measuring properly.

Advertising
This manual is related to the following products: