Use furnace temp for furnace setpoint, General setup, Security settings – Super Systems MGA6010 User Manual

Page 34: Analog output setup

Advertising
background image

MGA 6010 Operations Manual

Super Systems Inc.

Page 34 of 54

IMPORTANT!

This option will work only when the MGA 6010 is

not communicating with any slave instruments.

If you attempt to use this option by tapping it

and then tapping “Edit” when the MGA 6010 is

communicating with a slave instrument, the

message “Cannot use this feature with Slave

Instrument configured” will be displayed. To

disable slave instruments, open the

Communications and Source Setup

Atmosphere/Temp Sources menu; then set

both the Probe Temp/mV Instrument Address

and Furnace Temp Instrument Address to “0”.

Use Furnace Temp for Furnace Setpoint

When this option is set to “True”, the MGA 6010 will use the furnace temperature as the value used to

select an IR Shim Factor. When this option is set to “False”, the MGA 6010 will use the temperature

setpoint as the value used to select the IR Shim Factor.

General Setup

This screen shows the current time and date for the instrument, as well as the temperature scale. The

temperature mode can be changed from this screen, but the time cannot be changed unless the MGA

software is closed. This can be done through the following steps:

Manually Setting the Time

1.

Log in to the MGA 6010 touch screen using the supervisor access code (by default, this code is ‘2’).

2.

Access the Instrument Setup → General Setup menu.

3.

Select “Shut Down Software” followed by “Yes”.

4.

The program will shut down and a Windows CE screen will appear.

5.

Click the time in the lower right corner (press the stylus in this corner if the time does not

automatically appear).

6.

Set the time and date to the desired settings

7.

Select “OK”

8.

Double-click on the “SaveRegs” icon.

9.

Select “OK” from the Registry Save Complete notification.

10.

Turn the instrument off and wait for the screen to turn off.

11.

Turn the instrument back on.

12.

The instrument will automatically boot up as a gas analyzer with the IP address that has been

manually entered.

Security Settings

This page is used to change the pass code used for logging into the menu list. The default setting for the

Supervisor Pass Code is 1, and the default setting for Configuration is 2. The maximum value for either

code is 32767.

The Configuration Code will also work for all items on the Supervisor Pass Code items, so entering the

Configuration Code (default =2) will provide access to all available menus.

Analog Output Setup

The MGA 6010 has four analog outputs. These outputs can be configured for variable, zero value, and

span value. The Variable is the process variable that applies to that analog output. Possible values are

CO, CO

2

, H

2

, High CO

2

, IR % carbon, Gas ratio, and Gas squared ratio. The Zero Value is the value that

corresponds to 4mA on a 4-20mA scale. The Span Value is the value that corresponds to 20mA on a 4-

20mA scale.

Advertising
This manual is related to the following products: