cylinder. For example, the nominal value for CO may be 20%, but your cylinder may actually have 19.96%

CO. 19.96 is the value that should be entered as a Target.

After the gas values have been entered, proceed with the calibration in the same manner as with the zero

calibration. Never perform a span calibration without first doing a zero calibration.

Automatic Sampling Parameters

This instrument is capable of communicating directly with a control instrument. This is valuable because

it will provide real-time entry of the temperature and millivolt information from the probe, allowing for an

accurate comparison between the IR % Carbon and the Probe %Carbon. To establish this

communications link, see “Communications and Source Setup – Atmosphere/ Temp Sources.”

The Automatic Sampling Parameters screen will allow the user to adjust the way that the MGA 6010

updates the COF / PF in the atmosphere controller. All of the parameters on this page can be

disregarded if the “COF/PF Adjustment Mode” is set to Monitor mode. These parameters only apply when

the instrument is in Control mode.

COF/PF Adjustment Increment

When adjustments are made automatically, this value indicates the size of the step that is made when the

COF/PF is changed. It is recommended that this number remain low to avoid making sudden changes to

the process that could be caused by temporary conditions.

COF/PF Adjustment Interval (minutes)

This indicates the frequency that automatic adjustments are made. We recommend making small

changes at a frequent interval instead of making large changes at longer intervals. This will prevent

temporary changes in atmosphere from making dramatic adjustments to the COF/PF.