MGA 6010 Operations Manual

Super Systems Inc.

Page 42 of 54

3.

Prepare the temperature sourcing device. This device should be capable of sourcing a

temperature from type “S” and type “K” thermocouple types.

4.

Connect the thermocouple wire from the temperature sourcing device to the appropriate input on

the side of the case.

5.

Enter the desired temperature value based on which to deliver a corresponding voltage.

6.

Configure the temperature sourcing device to deliver the proper voltage.

7.

Wait for the “Current Value” displayed on the screen to get as close to the target voltage as

possible.

8.

Tap Calibrate. Tap “Yes” when asked if you want to proceed with the calibration. A progress

indicator will appear.

9.

When the process completes, temperature calibration is finished.

Valve Setup

The MGA 6010 uses four flow control valves for nitriding gas analysis. These valves are enabled and set

up in the Valve Setup window.