MGA 6010 Operations Manual

Super Systems Inc.

Page 19 of 54

Pump Control

The Pump Control screen will identify and allow the modification of the pump status (On or Off), as well as

the amount of time (in seconds) to delay turning the pump on (Pump On Delay) and off (Pump Off Delay).

Sensor Calibration

On the MGA 6010, you can perform a sensor calibration with one of four calibration sources. These

sources are displayed when Sensor Calibration is first selected: Automatic Calibration Port, Auxiliary

Calibration Port, Sample Line (with pump on), and Sample Line (with pump off).

You will be required to select one of the four sources.

With the Automatic Calibration Port, calibration is performed using the Zero Calibration Gas Inlet and

Span Calibration Gas Inlet ports on the side of the unit.