Factory Default Settings

Selecting this option will cause the instrument to revert back to the settings that it contained when it

came from Super Systems. Any changes or modifications made since then will be lost.

Other Settings

This screen displays calculated dew point. This value is not normally displayed because it is only

accurate during certain conditions. Calculating dew point by using the gas values requires assumptions

to be made regarding the composition of the gas being sampled. Since these assumptions are not always

accurate the resulting dew point is not always accurate. When using Endothermic gas in a heat treating

environment, the calculated dew point will usually be close to the correct value, but there is no substitute

for a dew point that is determined through direct measurement of the moisture content of the gas.

When the dew point is enabled, it will appear near the bottom left of the Carbon Calculation screen.

Language Setup

The instrument language should be pre-configured at Super Systems prior to shipment but can also be

changed by making a selection from the drop-down list. For the language change to take effect, the

instrument must be powered off and then back on again.