Gas Alarm Setup

The MGA 6010 allows the user to configure various alarms. For each parameter, the Lower Limit, Upper

Limit, and Action. As default, no alarms are enabled on the MGA unless a special request was made to do

so at the time of ordering. To modify a parameter, select the item and press Edit. A screen will appear

allow for the entry of the limits and the action. When any parameter is in an alarm state, a notification

banner will show at the top of the screen identifying the alarm, and the red light on the door will

illuminate. If desired, one of the two relays (or both simultaneously) can be energized. There are four

possible actions for the alarms:


None – On screen notification of alarm condition. No relay actions.

AL1 – On screen notification of alarm condition plus energizing of alarm relay #1.

AL2 – On screen notification of alarm condition plus energizing of alarm relay #2.

AL1 & AL2 – On screen notification of alarm condition plus energizing of alarm relays #1 and #2.

The relays provide a contact that can be connected to external lights, horns, or other devices as desired.

The rating for these relay contacts is maximum of 6 Amps / 250 Volts.

Instrument Information

These items cannot be modified; they can only be viewed.

General Information

This is information on the revision levels of various components of the instrument. This can be valuable

when consulting with Super Systems about issues with the instrument.