Maintain Sessions

The MGA 6010 is designed to delete the oldest files first in the event that the storage capacity has been

exceeded. It is also possible to delete old files manually by entering a date on this screen. All files that

were made before that date would be permanently deleted. It is important to note that after the data has

been downloaded to a computer, a copy is stored on that computer. The data will always be available on

the computer even if it has been deleted from the screen.

Maintain Users

The names of all potential users can be entered here. Names can be added and deleted as required.

Compact Database

Periodic database compaction will help make data storage more efficient and allow more data to be

stored before it is automatically deleted. Nothing is deleted or lost when this button is pressed, and the

only result will be a performance improvement.

Repair Database

Use this option to repair an existing database when and if you suspect there are errors to the existing

database.

Pressure Sensor Calibration

The pressure it set at Super Systems for local conditions. For optimal performance, the ambient

pressure should be reset at the final destination. This can be done by determining the barometric

pressure and the elevation and entering them on this screen. After the two values have been entered,

press the Calibrate button and the pressure sensor calibration will be complete.

Thermister Calibration

This will be set at Super Systems and should not need to be adjusted by the end user. It allows for the

sample gas temperature and the ambient temperature inside the instrument to be set. This should only

be performed after the instrument has been powered on long enough for it to achieve temperature

equilibrium.

SuperCalc

SuperCalc is a proprietary software tool developed by SSI to allow the user to perform different scenarios

and view the resulting percent carbon. It allows the user to enter gas percentages, probe information,

and temperatures to see the effects of each variable on the calculated percent carbon. The data on this

screen is independent of any values that are determined by the MGA 6010, and it is only provided as a

reference tool.

Set User Cal / Load User Cal

This feature allows the user to create new factory default calibration settings for the sensor. Instead of

reverting to the factory calibration values, it can revert back to different calibration settings. This is

accomplished by first setting the user calibration values. At any time after they are set, they can be

restored by selecting Load User Cal.

Analog Input Calibration

The True Temp MGA 6010 is equipped with an analog input card that is calibrated at the factory before the

True Temp unit is shipped. Optionally, you can verify the calibration or re-calibrate the unit at a later time

if desired. The Analog Input Calibration includes Zero and Span calibrations for millivolt input and

calibration of the actual thermocouple inputs based on the type of thermocouple wire used (the MGA 6010

supports calibration with type “K” and type “S” wire). This section provides more information on

performing those calibrations, if you wish to perform them. Note the warning below.