operating temperature. The controller can be configured to signal an alarm condition on either the control 
thermistor or the secondary thermistor or both. 

1.4

The TC‐48‐20 can be used with either one or two separate DC power supplies, depending primarily on the 
nominal operating voltage of the TE device. If the maximum TE device input voltage is less than 9 V, then 
two power supplies must be used with the controller. If the TE device can use a voltage ≥9 V but <50 V, then 
the controller can be used with either one or two power supplies.  

The controller can also be set up to provide linear control output when used with a power supply whose 
output  can  be  controlled  with  a  0  to  5  V  signal.  The  power  supply  can  then  provide  a  power  output 
proportional  to  the  signal  sent  by  the  controller.  This  setup  is  also  useful  for  situations  in  which  the  TE 
device  requires  more  power  than  what  the  controller  itself  can  directly  handle.  However,  this  setup  will 
reduce the overall control stability. The voltage signal from the controller is actually obtained by filtering the 
PWM  output, and  this  introduces a  slight delay  in  the control  response.  Also,  there will  be  a  delay  in  the 
response  of  the  power  supply  since  it  too  must  process  the  correct  power  output  based  on  the  sensed 
voltage input from the controller.  

When using one power supply for powering the controller and the TE device together, the power 
supply  input  voltage  is  passed  directly  through  the  controller  to  the  TE  device  during  the  “ON” 
pulse. The user should choose an input voltage that is required for the TE device and yet is also ≥9 
V but <50 V. The controller could be damaged if operated outside this voltage range. 

When using two power supplies, the controller input power supply must be ≥9 V but <50 V at 150 
mA minimum. The power supply input voltage for the TE device can be ≥0 V but <50 V. 

The  maximum  allowable  current  through  the  controller  is  20  A.  The  maximum  allowable  current 
draw for the TE device must therefore be less than 20 A depending on if and how much current is 
used to power alarm signals and the controller itself. The 20 A limit applies regardless of whether 
you are using one power supply or using two independent power supplies. 

The  controller  does  not  have  an  internal  fuse  or  circuitry  to  limit  current.  Therefore,  an  external 
fuse, appropriately sized for protecting the controller/TE device, should be connected between the 
power  supply  and  the  controller  to  prevent  damage  to  the  controller/TE  device  and  to  prevent 
injury to the user should an over‐current condition occur.  Alternately, a power supply with integral 
over  current  protection  can  be  used  if  it  is  appropriately  sized  for  protecting  the  controller/TE 
device. 

When  making  a  cooling  system  from  a  single  TE  device, the  maximum  operating  voltage  for  that 
system is usually no more than 75% of the rated Vmax of the TE module. The 75% rule is based on 
the TE module being thermally connected to a “good” heat sink; system modeling should be done 
to verify this rule is applicable though. If multiple TE modules are used in series or series‐parallel 
combination,  the  Vmax  of  the  system  will  be  approximately  75%  of  the  rated  Vmax  of  each  TE 
module multiplied by the number of modules in series. Applying a voltage greater than the system 
maximum  will  not  necessarily  damage  the  controller  (unless  voltage  and/or  current  limits  are 
exceeded), but the TE device could be damaged by overheating as a result. 

Power supply and TE Device wire leads should be kept as short as possible to minimize electrical 
losses and reduce the likelihood of generating unwanted electromagnetic interference. Use wires 
of  a  sufficient  gage  appropriate  to  the  amount  of  electrical  energy  each  wire  is  to  carry.  Wire 
insulation  and  size  must  also  be  appropriate  to  the  ambient  temperature  and/or  temperature  of 
objects  in  contact  with  the  wire.  Wire  leads  supplied  by  TE  Technology  are  for  prototyping 
purposes and should be reviewed for appropriateness in the final application. Wire length must not 
exceed one meter. However, see also warnings about wire length under section 1.6. 