The printed circuit board underneath the JP4 terminal block can reach a normal operating temperature of 
approximately 90 °C. The controller specifications are based on using wires connecting to JP4 to meet UL 
1015 requirements and have a wire size of 2.02 mm

2

 effective cross‐sectional area, which is comprised of 41 

strands of 0.254 mm diameter copper wire. The effective equivalent wire size is 14 AWG. The specifications 
are further based on a wire length of 410 mm between the power supply and the controller and a wire 
length of 920 mm between the controller and the TE device. Using wire with a smaller conductor cross 
section and/or shorter in length might cause abnormally high temperature to be present on the JP4 terminal 
block and wire. If smaller and/or shorter wire must be used, the amount of current the controller can safely 
accept might need to be decreased and/or the ambient temperature at which the controller can operate 
might need to be lowered. 

The wires inserted into connector JP4 should have a strip length of 7 mm, be fully inserted into the 
connector, and the JP4 screw terminals should be tightened with a minimum torque of 0.5 N‐m.  The wires 
inserted into connector JP5 should have a strip length of 5 mm, be fully inserted into the connector, and the 
JP5 screw terminals should be tightened with a minimum torque of 0.25 N‐m.  

 
1.7

Turn on power supply to the controller and power supply to the TE device if applicable and adjust controller 
settings as necessary. Controller tuning is discussed in section 3.  Remember, the factory default is set to 
have output power to the TE device enabled when the controller is powered. This can only be turned off by 
using the software.  Therefore, you should also be prepared to de‐energize the controller power in case 
there is an incorrectly set parameter which could cause a hazard. 

 

NOTE: The factory default assumes a secondary thermistor is connected to the controller. When not using a 
thermistor on the second sensor input the user will need to adjust the ALARM 2 FUNCTION to KEEP 
OUTPUT ON.
 See section 2.3.10 and 2.3.11 for further details. 

 

The keypad can reach a normal operating temperature of 70 °C. 

When  vertically  mounted  the  terminal  blocks  should  be  positioned  pointing  upwards,  with  the 
direction  of  gravity  parallel  to  the  plane  of  the  controller  mounting  surface.  The  maximum 
temperature specifications (vertical orientation) are based on this positioning. When mounted in an 
orientation between  horizontal  and  vertical,  the  terminal  blocks  should  be  mounted  in  the upward 
most position (with the direction of gravity defined as point downward).  These positions allow heat 
to  rise  upward,  out  of  the  opening  of  the  box.    Orientations  differing  from  this  could  cause  the 
controller to overheat. 

Do  not  mount  the  controller  to  a  surface  which  is  exposed  to  a  source  of  heat,  such  as  from 
electronics, machinery, or solar radiation. 

Do not cover the controller with any object or otherwise restrict natural convection airflow around 
the controller. Doing so could cause the controller to overheat. 

Do not mount the controller to an insulating surface. Doing so could cause the controller to overheat. 

Do not operate the controller in such a manner as to cause the surface temperature of the case or 
internal temperature to reach 70 °C. Otherwise the controller might be damaged and there might be 
a risk of fire as a result. 

Do not allow the controller to be exposed to water (such as from dripping or leaking water lines or in 
which the controller is operating at or below the dew point temperature) 

Do not allow metallic dust/shavings to enter into the controller box. 