Rosemount 3300 series, Reference manual – Emerson Guided Wave Radar Level and Interface Transmitters 3300 User Manual

Page 38

Advertising
background image

Reference Manual

00809-0100-4811, Rev CA

February 2006

Rosemount 3300 Series

3-14

Figure 3-11. Threaded tank
connection.

1. For tank connections with BSP/G threads,

place a gasket on top of the tank flange, or
use a sealant on the threads of the tank
connection.

2. Lower the transmitter and probe into the

tank.

3. Screw the adapter into the process

connection.

4. Loosen the nut that connects the

transmitter housing to the probe slightly.

5. Rotate the transmitter housing so the

cable entries/display face the desired
direction.

6. Tighten the nut.

NOTE!
For adapters with NPT threads, pressure-tight
joints require a sealant.

Sealant on threads or
gasket (for BSP/G threads)

Nut

Tank connection

Probe

T

R

A

N

SM

ITTE

R

_

M

O

U

N

T_

TH

R

E

AD

.EPS