Rosemount 3300 series, Reference manual – Emerson Guided Wave Radar Level and Interface Transmitters 3300 User Manual

Page 41

Advertising
background image

Reference Manual

00809-0100-4811, Rev CA
February 2006

3-17

Rosemount 3300 Series

Coaxial

To cut a coaxial probe do the
following:

NOTE!
The HTHP coaxial probe must not be
cut in field.

1. Insert the centering piece.

(The centering piece is
delivered from factory and
should be used to prevent the
spacers centering the rod from
coming loose
).

2. Cut the tube to the desired length.

3. Move the centering piece.

4. Cut the rod inside the tube. Make

sure that the rod is fixed with the
centering piece while cutting.

• Pipes longer than

49 inches/1250 mm can be
shortened by as much as
23.6 inches/600 mm.

• Pipes shorter than

49 inches/1250 mm can be cut as
long as the remaining length is not
less than
15.7 inches/400 mm.

5. Update the transmitter

configuration to the new probe
length, see “Probe Length“ on
page 4-9.

CO

AX

IA

L

_

CUT.T

IF

Centering piece

Maximum shortening
23.6 inches / 600 mm

L>49 inches/
1250 mm

CO

AXI

AL SHO

R

T1

Minimum probe length
15.7 inches / 400 mm

L

49 inches/

1250 mm

CO

AXI

AL SHO

R

T2