Reference Manual

00809-0100-4811, Rev CA
February 2006

4-17

Rosemount 3300 Series

Using the Setup
Function

To install a 3300 transmitter by using the Setup function do the following:

Figure 4-11. RCT workspace

Figure 4-12. Setup Info

NOTE
When working with the Setup window keep in mind that for all tabs except the
Info tab, data is updated by clicking the Receive button. To download data to
the transmitter click the Send button.

1. Start the RCT software.

2. In the RCT workspace click the

Setup icon (make sure that the
Basic area is open),
or
choose the View>Setup menu
option.

Setup

Basic

RCT

-RCT1

3. Choose the appropriate tab:

Info: information about the device.

Basics: Set Probe Type and
measurement units.

Analog: Variable assignment and
range value settings.

Tank Config: Tank height and other
geometry settings, dielectric
constants for vapor and upper
product.

Volume: specification of tank
geometry for volume calculations.

LCD: display panel settings.

R

C

T

-SETUP_

INFO

Advertising